Bump Testing GX-2012s and Gas Tracers 192
calibration will automatically begin on the %volume sensors for those
instruments and GAS 2 calibration gas will continue to flow.
22. After GAS 2 calibration gas has been applied, the docking station will purge
with fresh air for 45 seconds.
Figure 152: Fresh Air Purge
Vedere la pagina 196