33 M2 Rig Monitor Operator’s Manual
9. Reinstall the junction box cover.
10. Connect the battery clamps to the + and - terminals of the 12 VDC battery.
CAUTION: Allow the replacement detector to warm up for 5 minutes before you continue with
the next step.
11. Calibrate the replacement detector as described in the Calibration section of this
manual.