38 • Setting Commands
When you upload a parameter configuration file to an
instrument, you are uploading instrument parameters such as
alarm points, alarm pattern, calibration frequency, etc.
Uploading a parameter configuration file to an instrument will
not change the instrument identification parameters, serial
number, user ID, and station ID, if the file was created from a
different instrument.
Follow these steps to load a parameter configuration file to the
program and upload it to an instrument:
1. Launch the GX-2009 User Setup Program as described in
“Launching the Program” on page 14.
2. Connect the GX-2009 to the User Setup Program as
described in “Connecting a GX-2009 to the User Setup
Program” on page 15.
NOTE: It is not necessary to use Instrument Information or
Instrument Get & Save in order to load a parameter
configuration file and upload it to an instrument.
Although it is not necessary to use Instrument
Information or Instrument Get & Save before loading
a file and uploading it to an instrument, you can do so
after using either of these functions if desired.