Model GX-2001 Operator’s Manual Maintenance 36
Replacing the Sensor Cover
WARNING: Replace the sensor cover in a non-hazardous
environment.
1. Verify that the Model GX-2001 is off.
2. Unscrew the two screws that secure the sensor retainer and sensor
cover retainer to the Model GX-2001 and remove them.
3. Using a small flat blade screwdriver, gently pry each side of the sensor
cover retainer away from the sensor retainer.
4. Remove the sensor cover, then insert the replacement cover.
5. Reattach the sensor cover retainer in its original position. It should snap
into place onto the sensor retainer.
6. Replace the two screws that secure the sensor retainer and sensor
cover retainer to the Model GX-2001, which you removed in step 2
above.